- Medical masks
- Protective face coverings
- Coated fabrics
- Barrier materials
Synthetic Blood Penetration Tester
Evaluate resistance to synthetic-blood penetration under defined challenge conditions.
A synthetic-blood penetration tester for medical masks and other coated protective materials under defined challenge pressure.

Workflow and key features
The CTY-01A projects a specified volume of synthetic blood at the specimen and supports visual evaluation of penetration. Controlled pressure and geometry help compare barrier performance relevant to medical protective products.
- Defined synthetic-blood challenge replaces uncontrolled splash tests
- Adjustable challenge pressure supports different mask classifications
- Applicable to medical masks and configured coated protective materials
- Consistent target distance and specimen mounting improve repeatability
- Supports product development, type testing and routine quality checks
Reference configuration
Published values describe the referenced standard configuration. The formal quotation confirms the selected range, fixtures, utilities, method edition and supplied documentation.
- Required air supply
- 0.7-1.0 MPa through a 6 mm polyurethane tube
- Pressure vessel adjustment
- 0-40 kPa
- Dispensing-valve adjustment
- 0-40 kPa
- Timing resolution
- 0.01 s
- Test medium
- Air
- Control
- Microcomputer
- Net weight
- 12 kg
- Electrical supply
- 220 VAC, 50 Hz
Test-method referencesWorkflow references, not a certification claim4 references +
These references identify workflows described for this model. The formal quotation confirms the exact edition, fixtures, software, accessories and verification documents supplied for the selected configuration.
- YY/T 0969-2013
- GB 19083-2010
- YY/T 0691-2008
- ISO 22609
- Device or protective-article type
- Required test method
- Challenge conditions or expected force range
- Sample dimensions and quantity
The quoted configuration remains the controlling specification.
Published values describe the referenced standard configuration. The formal quotation confirms the selected range, fixtures, utilities, method edition and supplied documentation. Delivery scope, warranty, lead time, calibration evidence and overseas service are stated in the formal offer.